MEASURING AN AC VOLTAGE
In oscilloscope measurements, the AC component of a
waveform is usually measured in terms of its peak-to-
peak value. You can easily make this type of measure-
ment by using the grid overlay on the oscilloscope
screen to measure the vertical distance between peaks
and multiplying this distance by the oscilloscope
screen’s deflection factor. The result is the actual peak-
to-peak voltage.
Follow these steps to measure a waveform’s peak-to-
peak AC component.
Caution
: To avoid an overload when measuring AC,
make sure the zero line is centered vertically on the
grid overlay.
1. Set the input coupling switch to
GND
.
Note
: When you set the input coupling switch to
GND
, the ProbeScope automatically resets the
trigger mode to
AUTO
.
2. Connect the
GROUND
(0 volt) wire to the signal
source’s ground or reference level.
3. Rotate
Zero Level
until the zero reference line is
centered vertically on the grid overlay.
Note
: If the trace is widened by stray interference,
ground the probe body near the point being mea-
sured. The trace position is the voltage reference
line and all voltage measurements are read in
respect to this line. (Do not adjust the vertical
positioning control after the reference is estab-
lished.)